HQ:NSC15/Cr-Au BS
Gold Coated Tapping Mode AFM Probe
AFM probes of the HQ:NSC15 series are generally used in tapping mode for imaging hard samples when high topographic and phase contrast are necessary. These AFM probes are also suitable for non-contact mode.
The HQ AFM probes offer high consistency of the AFM tip radius, the AFM cantilever reflectivity and the quality factor.
The gold reflective coating enhances the laser reflectivity of the AFM cantilever in air and liquids.
Coating
Reflective Gold
AFM Probe Specifications
AFM Tip
Shape | Height | Full Cone Angle | Radius |
---|---|---|---|
Shape Rotated | Height 15 µm (12 - 18 µm)* | Full Cone Angle 40° | Radius < 8 nm |
AFM Cantilever
Cantilever | Shape | Force Const. | Res. Freq. | Length | Width | Thickness |
---|---|---|---|---|---|---|
Cantilever Cantilever A | Shape Beam | Force Const. 40 N/m(20 - 80 N/m)* | Res. Freq. 325 kHz(265 - 410 kHz)* | Length 125 µm(1 - 130µm)* | Width 30 µm(27 - 33µm)* | Thickness 4µm(3.5 - 4.5 µm)* |
* typical values
Videos